An in-vitro biomechanical comparison of the orosco and AO locking plates for anterior cervical spine fixation

S. A. Smith, Ronald Lindsey, B. J. Doherty, J. Alexander, J. H. Dickson

Research output: Contribution to journalArticle

28 Citations (Scopus)

Abstract

The Prosco anterior cervical spine plate (H-plate) and the Morscher AO cervical locking plate (CSLP) were studied to determine their comparative in vitro mechanical properties in flexion. Human cadaver cervical spines were tested nondestructively in flexion as intact, destabilized, and stabilized specimens with both implants. Stabilized specimens were also subjected to a large angular displacement in an attempt to induce implant failure. Differences between intact, destabilized, and stabilized specimens were significant for both the CSLP and H-plate groups in flexion testing (p < 0.05). Large angular displacement testing resulted in plate/screw displacement with the H-plate but not with the CSLP. Small angular displacements in flexion was tolerated well by both implants, but only the CSLP maintained stability with large angular displacements.

Original languageEnglish (US)
Pages (from-to)220-223
Number of pages4
JournalJournal of Spinal Disorders
Volume8
Issue number3
DOIs
StatePublished - 1995
Externally publishedYes

Fingerprint

Spine
Cadaver
In Vitro Techniques

Keywords

  • Anterior fixation
  • Biomechanics
  • Cervical spine
  • Plate
  • Stability

ASJC Scopus subject areas

  • Clinical Neurology
  • Surgery

Cite this

An in-vitro biomechanical comparison of the orosco and AO locking plates for anterior cervical spine fixation. / Smith, S. A.; Lindsey, Ronald; Doherty, B. J.; Alexander, J.; Dickson, J. H.

In: Journal of Spinal Disorders, Vol. 8, No. 3, 1995, p. 220-223.

Research output: Contribution to journalArticle

Smith, S. A. ; Lindsey, Ronald ; Doherty, B. J. ; Alexander, J. ; Dickson, J. H. / An in-vitro biomechanical comparison of the orosco and AO locking plates for anterior cervical spine fixation. In: Journal of Spinal Disorders. 1995 ; Vol. 8, No. 3. pp. 220-223.
@article{0a743fb300b74bd4ba2ae69f82c09d40,
title = "An in-vitro biomechanical comparison of the orosco and AO locking plates for anterior cervical spine fixation",
abstract = "The Prosco anterior cervical spine plate (H-plate) and the Morscher AO cervical locking plate (CSLP) were studied to determine their comparative in vitro mechanical properties in flexion. Human cadaver cervical spines were tested nondestructively in flexion as intact, destabilized, and stabilized specimens with both implants. Stabilized specimens were also subjected to a large angular displacement in an attempt to induce implant failure. Differences between intact, destabilized, and stabilized specimens were significant for both the CSLP and H-plate groups in flexion testing (p < 0.05). Large angular displacement testing resulted in plate/screw displacement with the H-plate but not with the CSLP. Small angular displacements in flexion was tolerated well by both implants, but only the CSLP maintained stability with large angular displacements.",
keywords = "Anterior fixation, Biomechanics, Cervical spine, Plate, Stability",
author = "Smith, {S. A.} and Ronald Lindsey and Doherty, {B. J.} and J. Alexander and Dickson, {J. H.}",
year = "1995",
doi = "10.1097/00002517-199506000-00007",
language = "English (US)",
volume = "8",
pages = "220--223",
journal = "Journal of Spinal Disorders",
issn = "1536-0652",
publisher = "Lippincott Williams and Wilkins",
number = "3",

}

TY - JOUR

T1 - An in-vitro biomechanical comparison of the orosco and AO locking plates for anterior cervical spine fixation

AU - Smith, S. A.

AU - Lindsey, Ronald

AU - Doherty, B. J.

AU - Alexander, J.

AU - Dickson, J. H.

PY - 1995

Y1 - 1995

N2 - The Prosco anterior cervical spine plate (H-plate) and the Morscher AO cervical locking plate (CSLP) were studied to determine their comparative in vitro mechanical properties in flexion. Human cadaver cervical spines were tested nondestructively in flexion as intact, destabilized, and stabilized specimens with both implants. Stabilized specimens were also subjected to a large angular displacement in an attempt to induce implant failure. Differences between intact, destabilized, and stabilized specimens were significant for both the CSLP and H-plate groups in flexion testing (p < 0.05). Large angular displacement testing resulted in plate/screw displacement with the H-plate but not with the CSLP. Small angular displacements in flexion was tolerated well by both implants, but only the CSLP maintained stability with large angular displacements.

AB - The Prosco anterior cervical spine plate (H-plate) and the Morscher AO cervical locking plate (CSLP) were studied to determine their comparative in vitro mechanical properties in flexion. Human cadaver cervical spines were tested nondestructively in flexion as intact, destabilized, and stabilized specimens with both implants. Stabilized specimens were also subjected to a large angular displacement in an attempt to induce implant failure. Differences between intact, destabilized, and stabilized specimens were significant for both the CSLP and H-plate groups in flexion testing (p < 0.05). Large angular displacement testing resulted in plate/screw displacement with the H-plate but not with the CSLP. Small angular displacements in flexion was tolerated well by both implants, but only the CSLP maintained stability with large angular displacements.

KW - Anterior fixation

KW - Biomechanics

KW - Cervical spine

KW - Plate

KW - Stability

UR - http://www.scopus.com/inward/record.url?scp=0029063891&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029063891&partnerID=8YFLogxK

U2 - 10.1097/00002517-199506000-00007

DO - 10.1097/00002517-199506000-00007

M3 - Article

VL - 8

SP - 220

EP - 223

JO - Journal of Spinal Disorders

JF - Journal of Spinal Disorders

SN - 1536-0652

IS - 3

ER -