Bending of the Cotrel-Dubousset instrumentation after direct trauma

A case report

Arvind Nana, Zbigniew Gugala, Ronald Lindsey, Pedro M. Caram, Jesse H. Dickson

Research output: Contribution to journalArticle

2 Citations (Scopus)

Abstract

Study Design. Case report. Objectives. To describe a fracture through the fusion mass of a spine that had been corrected previously with Cotrel- Dubousset rods. These rods had failed in bending after direct trauma. Summary of Background Data. Nine years after successful treatment of scoliosis with Cotrel-Dubousset instrumentation, the patient was in a motor vehicle accident and sustained a hyperextension spine injury with complete L1-L2 paraplegia and disruption of the fusion mass. The Cotrel-Dubousset instrumentation rods, which failed in bending, could not be corrected in situ, and the angulated segments had to be resected. The spine then became extremely unstable, and the patient consulted the authors for definitive stabilization. Results. The spine was stabilized by attaching the proximal and distal retained Cotrel- Dubousset instrumentation to supplemental rods in a 'domino' fashion. Crosslinks were added to improve the torsional stability. Intraoperatively, the fracture was well reduced, and the fixation was stable. A posterolateral fusion was performed with allogenic bone graft. Conclusion. Bent Cotrel- Dubousset instrumentation rods are still very strong and may not correct in situ. If resection is required, the retained portions of Cotrel-Dubousset instrumentation can serve as attachments to restore stable fixation a 'domino' technique.

Original languageEnglish (US)
Pages (from-to)891-894
Number of pages4
JournalSpine
Volume25
Issue number7
DOIs
StatePublished - Apr 1 2000
Externally publishedYes

Fingerprint

Spine
Wounds and Injuries
Paraplegia
Scoliosis
Motor Vehicles
Accidents
Transplants
Bone and Bones
Therapeutics

Keywords

  • Cotrel-Dubousset
  • Hardware failure
  • Instrumentation
  • Spine fracture

ASJC Scopus subject areas

  • Physiology
  • Clinical Neurology
  • Orthopedics and Sports Medicine

Cite this

Bending of the Cotrel-Dubousset instrumentation after direct trauma : A case report. / Nana, Arvind; Gugala, Zbigniew; Lindsey, Ronald; Caram, Pedro M.; Dickson, Jesse H.

In: Spine, Vol. 25, No. 7, 01.04.2000, p. 891-894.

Research output: Contribution to journalArticle

Nana, Arvind ; Gugala, Zbigniew ; Lindsey, Ronald ; Caram, Pedro M. ; Dickson, Jesse H. / Bending of the Cotrel-Dubousset instrumentation after direct trauma : A case report. In: Spine. 2000 ; Vol. 25, No. 7. pp. 891-894.
@article{50d381fe55844b46aab080ca9d6d1f06,
title = "Bending of the Cotrel-Dubousset instrumentation after direct trauma: A case report",
abstract = "Study Design. Case report. Objectives. To describe a fracture through the fusion mass of a spine that had been corrected previously with Cotrel- Dubousset rods. These rods had failed in bending after direct trauma. Summary of Background Data. Nine years after successful treatment of scoliosis with Cotrel-Dubousset instrumentation, the patient was in a motor vehicle accident and sustained a hyperextension spine injury with complete L1-L2 paraplegia and disruption of the fusion mass. The Cotrel-Dubousset instrumentation rods, which failed in bending, could not be corrected in situ, and the angulated segments had to be resected. The spine then became extremely unstable, and the patient consulted the authors for definitive stabilization. Results. The spine was stabilized by attaching the proximal and distal retained Cotrel- Dubousset instrumentation to supplemental rods in a 'domino' fashion. Crosslinks were added to improve the torsional stability. Intraoperatively, the fracture was well reduced, and the fixation was stable. A posterolateral fusion was performed with allogenic bone graft. Conclusion. Bent Cotrel- Dubousset instrumentation rods are still very strong and may not correct in situ. If resection is required, the retained portions of Cotrel-Dubousset instrumentation can serve as attachments to restore stable fixation a 'domino' technique.",
keywords = "Cotrel-Dubousset, Hardware failure, Instrumentation, Spine fracture",
author = "Arvind Nana and Zbigniew Gugala and Ronald Lindsey and Caram, {Pedro M.} and Dickson, {Jesse H.}",
year = "2000",
month = "4",
day = "1",
doi = "10.1097/00007632-200004010-00021",
language = "English (US)",
volume = "25",
pages = "891--894",
journal = "Spine",
issn = "0362-2436",
publisher = "Lippincott Williams and Wilkins",
number = "7",

}

TY - JOUR

T1 - Bending of the Cotrel-Dubousset instrumentation after direct trauma

T2 - A case report

AU - Nana, Arvind

AU - Gugala, Zbigniew

AU - Lindsey, Ronald

AU - Caram, Pedro M.

AU - Dickson, Jesse H.

PY - 2000/4/1

Y1 - 2000/4/1

N2 - Study Design. Case report. Objectives. To describe a fracture through the fusion mass of a spine that had been corrected previously with Cotrel- Dubousset rods. These rods had failed in bending after direct trauma. Summary of Background Data. Nine years after successful treatment of scoliosis with Cotrel-Dubousset instrumentation, the patient was in a motor vehicle accident and sustained a hyperextension spine injury with complete L1-L2 paraplegia and disruption of the fusion mass. The Cotrel-Dubousset instrumentation rods, which failed in bending, could not be corrected in situ, and the angulated segments had to be resected. The spine then became extremely unstable, and the patient consulted the authors for definitive stabilization. Results. The spine was stabilized by attaching the proximal and distal retained Cotrel- Dubousset instrumentation to supplemental rods in a 'domino' fashion. Crosslinks were added to improve the torsional stability. Intraoperatively, the fracture was well reduced, and the fixation was stable. A posterolateral fusion was performed with allogenic bone graft. Conclusion. Bent Cotrel- Dubousset instrumentation rods are still very strong and may not correct in situ. If resection is required, the retained portions of Cotrel-Dubousset instrumentation can serve as attachments to restore stable fixation a 'domino' technique.

AB - Study Design. Case report. Objectives. To describe a fracture through the fusion mass of a spine that had been corrected previously with Cotrel- Dubousset rods. These rods had failed in bending after direct trauma. Summary of Background Data. Nine years after successful treatment of scoliosis with Cotrel-Dubousset instrumentation, the patient was in a motor vehicle accident and sustained a hyperextension spine injury with complete L1-L2 paraplegia and disruption of the fusion mass. The Cotrel-Dubousset instrumentation rods, which failed in bending, could not be corrected in situ, and the angulated segments had to be resected. The spine then became extremely unstable, and the patient consulted the authors for definitive stabilization. Results. The spine was stabilized by attaching the proximal and distal retained Cotrel- Dubousset instrumentation to supplemental rods in a 'domino' fashion. Crosslinks were added to improve the torsional stability. Intraoperatively, the fracture was well reduced, and the fixation was stable. A posterolateral fusion was performed with allogenic bone graft. Conclusion. Bent Cotrel- Dubousset instrumentation rods are still very strong and may not correct in situ. If resection is required, the retained portions of Cotrel-Dubousset instrumentation can serve as attachments to restore stable fixation a 'domino' technique.

KW - Cotrel-Dubousset

KW - Hardware failure

KW - Instrumentation

KW - Spine fracture

UR - http://www.scopus.com/inward/record.url?scp=0034003535&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0034003535&partnerID=8YFLogxK

U2 - 10.1097/00007632-200004010-00021

DO - 10.1097/00007632-200004010-00021

M3 - Article

VL - 25

SP - 891

EP - 894

JO - Spine

JF - Spine

SN - 0362-2436

IS - 7

ER -