BACKGROUND: If a self-expanding stent has been placed during endovascular treatment of an aneurysm and subsequently an open aneurysm surgery becomes necessary in the same or an adjacent area, is it possible and safe to obtain proximal control by placing a temporary clip on the artery at a point where it contains the stent? OBJECTIVE: To evaluate the effect of temporary clip application to 3 separate stent systems in an in vitro flow model with the stated hypothesis that clip application to these stents will result in permanent stent deformation. METHODS: This is an in vitro flow model study using an accepted synthetic blood vessel substitute. The Neuroform (Boston Scientific), Enterprise (Cordis/Codman), and Pipeline (ev3) stents were deployed within the flow model; temporary clips were applied; and angiographic measurements subsequently made. RESULTS: Two 4 × 30-mm Neuroform stents, two 4.5 × 28-mm Enterprise stents, and two 3.75 × 20-mm Pipeline stents were successfully deployed and clipped repeatedly (4 iterations). Two- and 3-dimensional angiograms were obtained. After repeated clip occlusion, the Neuroform and Enterprise stents returned to their original configuration and diameter. Clip application to both also resulted in immediate flow arrest. In contrast, initial clip application to the Pipeline stents did not result in flow arrest, but the second single clip application did. The Pipeline stents were also irreversibly deformed after the experimental protocol, with an average luminal diameter reduction of 26.85% (P <.05). CONCLUSION: The Neuroform and Enterprise stents responded favorably to temporary clip application, returning to their original diameter after clip removal and showing no sign of permanent structural modification. The Pipeline flow-diverting stent, however, was irreversibly deformed by clip application. These data indicate that temporary clip application to certain stents is possible. Further in vivo study is required.

Original languageEnglish (US)
Pages (from-to)1056-1062
Number of pages7
JournalNeurosurgery
Volume68
Issue number4
DOIs
StatePublished - Apr 2011

Fingerprint

Stents
Aneurysm
Blood Vessels
Emergencies
Surgical Instruments
Neurosurgeons
Blood Substitutes
Angiography
Arteries

Keywords

  • Aneurysm clip
  • Aneurysm stent
  • Coil embolization
  • Endovascular

ASJC Scopus subject areas

  • Clinical Neurology
  • Surgery

Cite this

Can a self-expanding aneurysm stent be clipped? Emergency proximal control options for the vascular neurosurgeon. / Bell, Randy S.; Bank, William O.; Armonda, Rocco A.; Vo, Alexander; Kerber, Charles W.

In: Neurosurgery, Vol. 68, No. 4, 04.2011, p. 1056-1062.

Research output: Contribution to journalArticle

Bell, Randy S. ; Bank, William O. ; Armonda, Rocco A. ; Vo, Alexander ; Kerber, Charles W. / Can a self-expanding aneurysm stent be clipped? Emergency proximal control options for the vascular neurosurgeon. In: Neurosurgery. 2011 ; Vol. 68, No. 4. pp. 1056-1062.
@article{672f17de79934e5a810c6025fdc607a9,
title = "Can a self-expanding aneurysm stent be clipped? Emergency proximal control options for the vascular neurosurgeon",
abstract = "BACKGROUND: If a self-expanding stent has been placed during endovascular treatment of an aneurysm and subsequently an open aneurysm surgery becomes necessary in the same or an adjacent area, is it possible and safe to obtain proximal control by placing a temporary clip on the artery at a point where it contains the stent? OBJECTIVE: To evaluate the effect of temporary clip application to 3 separate stent systems in an in vitro flow model with the stated hypothesis that clip application to these stents will result in permanent stent deformation. METHODS: This is an in vitro flow model study using an accepted synthetic blood vessel substitute. The Neuroform (Boston Scientific), Enterprise (Cordis/Codman), and Pipeline (ev3) stents were deployed within the flow model; temporary clips were applied; and angiographic measurements subsequently made. RESULTS: Two 4 × 30-mm Neuroform stents, two 4.5 × 28-mm Enterprise stents, and two 3.75 × 20-mm Pipeline stents were successfully deployed and clipped repeatedly (4 iterations). Two- and 3-dimensional angiograms were obtained. After repeated clip occlusion, the Neuroform and Enterprise stents returned to their original configuration and diameter. Clip application to both also resulted in immediate flow arrest. In contrast, initial clip application to the Pipeline stents did not result in flow arrest, but the second single clip application did. The Pipeline stents were also irreversibly deformed after the experimental protocol, with an average luminal diameter reduction of 26.85{\%} (P <.05). CONCLUSION: The Neuroform and Enterprise stents responded favorably to temporary clip application, returning to their original diameter after clip removal and showing no sign of permanent structural modification. The Pipeline flow-diverting stent, however, was irreversibly deformed by clip application. These data indicate that temporary clip application to certain stents is possible. Further in vivo study is required.",
keywords = "Aneurysm clip, Aneurysm stent, Coil embolization, Endovascular",
author = "Bell, {Randy S.} and Bank, {William O.} and Armonda, {Rocco A.} and Alexander Vo and Kerber, {Charles W.}",
year = "2011",
month = "4",
doi = "10.1227/NEU.0b013e31820d5396",
language = "English (US)",
volume = "68",
pages = "1056--1062",
journal = "Neurosurgery",
issn = "0148-396X",
publisher = "Lippincott Williams and Wilkins",
number = "4",

}

TY - JOUR

T1 - Can a self-expanding aneurysm stent be clipped? Emergency proximal control options for the vascular neurosurgeon

AU - Bell, Randy S.

AU - Bank, William O.

AU - Armonda, Rocco A.

AU - Vo, Alexander

AU - Kerber, Charles W.

PY - 2011/4

Y1 - 2011/4

N2 - BACKGROUND: If a self-expanding stent has been placed during endovascular treatment of an aneurysm and subsequently an open aneurysm surgery becomes necessary in the same or an adjacent area, is it possible and safe to obtain proximal control by placing a temporary clip on the artery at a point where it contains the stent? OBJECTIVE: To evaluate the effect of temporary clip application to 3 separate stent systems in an in vitro flow model with the stated hypothesis that clip application to these stents will result in permanent stent deformation. METHODS: This is an in vitro flow model study using an accepted synthetic blood vessel substitute. The Neuroform (Boston Scientific), Enterprise (Cordis/Codman), and Pipeline (ev3) stents were deployed within the flow model; temporary clips were applied; and angiographic measurements subsequently made. RESULTS: Two 4 × 30-mm Neuroform stents, two 4.5 × 28-mm Enterprise stents, and two 3.75 × 20-mm Pipeline stents were successfully deployed and clipped repeatedly (4 iterations). Two- and 3-dimensional angiograms were obtained. After repeated clip occlusion, the Neuroform and Enterprise stents returned to their original configuration and diameter. Clip application to both also resulted in immediate flow arrest. In contrast, initial clip application to the Pipeline stents did not result in flow arrest, but the second single clip application did. The Pipeline stents were also irreversibly deformed after the experimental protocol, with an average luminal diameter reduction of 26.85% (P <.05). CONCLUSION: The Neuroform and Enterprise stents responded favorably to temporary clip application, returning to their original diameter after clip removal and showing no sign of permanent structural modification. The Pipeline flow-diverting stent, however, was irreversibly deformed by clip application. These data indicate that temporary clip application to certain stents is possible. Further in vivo study is required.

AB - BACKGROUND: If a self-expanding stent has been placed during endovascular treatment of an aneurysm and subsequently an open aneurysm surgery becomes necessary in the same or an adjacent area, is it possible and safe to obtain proximal control by placing a temporary clip on the artery at a point where it contains the stent? OBJECTIVE: To evaluate the effect of temporary clip application to 3 separate stent systems in an in vitro flow model with the stated hypothesis that clip application to these stents will result in permanent stent deformation. METHODS: This is an in vitro flow model study using an accepted synthetic blood vessel substitute. The Neuroform (Boston Scientific), Enterprise (Cordis/Codman), and Pipeline (ev3) stents were deployed within the flow model; temporary clips were applied; and angiographic measurements subsequently made. RESULTS: Two 4 × 30-mm Neuroform stents, two 4.5 × 28-mm Enterprise stents, and two 3.75 × 20-mm Pipeline stents were successfully deployed and clipped repeatedly (4 iterations). Two- and 3-dimensional angiograms were obtained. After repeated clip occlusion, the Neuroform and Enterprise stents returned to their original configuration and diameter. Clip application to both also resulted in immediate flow arrest. In contrast, initial clip application to the Pipeline stents did not result in flow arrest, but the second single clip application did. The Pipeline stents were also irreversibly deformed after the experimental protocol, with an average luminal diameter reduction of 26.85% (P <.05). CONCLUSION: The Neuroform and Enterprise stents responded favorably to temporary clip application, returning to their original diameter after clip removal and showing no sign of permanent structural modification. The Pipeline flow-diverting stent, however, was irreversibly deformed by clip application. These data indicate that temporary clip application to certain stents is possible. Further in vivo study is required.

KW - Aneurysm clip

KW - Aneurysm stent

KW - Coil embolization

KW - Endovascular

UR - http://www.scopus.com/inward/record.url?scp=79952900702&partnerID=8YFLogxK