Cervical spine locking plate

in vitro biomechanical testing

Stephen A. Smith, Ronald Lindsey, Brian J. Doherty, Jerry W. Alexander, Jessie H. Dickson

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

The AO cervical spine locking plate (CSLP) for anterior subaxial fixation was recently received increasing clinical acclaim, yet to date the in vitro mechanical properties of this implant have not been reported. To determine the in vitro biomechanical properties of this device, five fresh human cadaver cervical spines were subjected to nondestructive testing in flexion and torsion in three stages: stage 1: intact spine; stage 2: destabilized spine; stage 3: destabilized spine with CSLP. Stage 3 specimens were also subjected to large angular displacement testing to assess the integrity of the fixation. In flexion, mean spinous process displacement was 1.21 mm for stage 1, 3.19 mm for stage 2, and 1.37 mm, for stage 3. Mean torsional stiffness was 2.86 Nm/degree in stage 1, 1.82 Nm/degree in stage 2, and 2.20 Nm/degree in stage 3. Large angular displacement testing in stage 3 resulted in screw loosening from the bone in two specimens; no screw plate loosening occurred. In our severely destabilized in vitro model, the CSLP restored flexion stability but not rotational stability. This suggests that supplemented bracing or fixation may be required to restore torsional stability.

Original languageEnglish (US)
Pages (from-to)222-225
Number of pages4
JournalEuropean Spine Journal
Volume1
Issue number4
DOIs
StatePublished - Mar 1993
Externally publishedYes

Fingerprint

Spine
Bone Screws
In Vitro Techniques
Cadaver
Equipment and Supplies

Keywords

  • Anterior fixation
  • Biomechanics
  • Cervical spine
  • Plate
  • Stability

ASJC Scopus subject areas

  • Surgery
  • Orthopedics and Sports Medicine

Cite this

Smith, S. A., Lindsey, R., Doherty, B. J., Alexander, J. W., & Dickson, J. H. (1993). Cervical spine locking plate: in vitro biomechanical testing. European Spine Journal, 1(4), 222-225. https://doi.org/10.1007/BF00298363

Cervical spine locking plate : in vitro biomechanical testing. / Smith, Stephen A.; Lindsey, Ronald; Doherty, Brian J.; Alexander, Jerry W.; Dickson, Jessie H.

In: European Spine Journal, Vol. 1, No. 4, 03.1993, p. 222-225.

Research output: Contribution to journalArticle

Smith, SA, Lindsey, R, Doherty, BJ, Alexander, JW & Dickson, JH 1993, 'Cervical spine locking plate: in vitro biomechanical testing', European Spine Journal, vol. 1, no. 4, pp. 222-225. https://doi.org/10.1007/BF00298363
Smith, Stephen A. ; Lindsey, Ronald ; Doherty, Brian J. ; Alexander, Jerry W. ; Dickson, Jessie H. / Cervical spine locking plate : in vitro biomechanical testing. In: European Spine Journal. 1993 ; Vol. 1, No. 4. pp. 222-225.
@article{7a03b7b52c134234ad57b2c26f5b3eb0,
title = "Cervical spine locking plate: in vitro biomechanical testing",
abstract = "The AO cervical spine locking plate (CSLP) for anterior subaxial fixation was recently received increasing clinical acclaim, yet to date the in vitro mechanical properties of this implant have not been reported. To determine the in vitro biomechanical properties of this device, five fresh human cadaver cervical spines were subjected to nondestructive testing in flexion and torsion in three stages: stage 1: intact spine; stage 2: destabilized spine; stage 3: destabilized spine with CSLP. Stage 3 specimens were also subjected to large angular displacement testing to assess the integrity of the fixation. In flexion, mean spinous process displacement was 1.21 mm for stage 1, 3.19 mm for stage 2, and 1.37 mm, for stage 3. Mean torsional stiffness was 2.86 Nm/degree in stage 1, 1.82 Nm/degree in stage 2, and 2.20 Nm/degree in stage 3. Large angular displacement testing in stage 3 resulted in screw loosening from the bone in two specimens; no screw plate loosening occurred. In our severely destabilized in vitro model, the CSLP restored flexion stability but not rotational stability. This suggests that supplemented bracing or fixation may be required to restore torsional stability.",
keywords = "Anterior fixation, Biomechanics, Cervical spine, Plate, Stability",
author = "Smith, {Stephen A.} and Ronald Lindsey and Doherty, {Brian J.} and Alexander, {Jerry W.} and Dickson, {Jessie H.}",
year = "1993",
month = "3",
doi = "10.1007/BF00298363",
language = "English (US)",
volume = "1",
pages = "222--225",
journal = "European Spine Journal",
issn = "0940-6719",
publisher = "Springer Verlag",
number = "4",

}

TY - JOUR

T1 - Cervical spine locking plate

T2 - in vitro biomechanical testing

AU - Smith, Stephen A.

AU - Lindsey, Ronald

AU - Doherty, Brian J.

AU - Alexander, Jerry W.

AU - Dickson, Jessie H.

PY - 1993/3

Y1 - 1993/3

N2 - The AO cervical spine locking plate (CSLP) for anterior subaxial fixation was recently received increasing clinical acclaim, yet to date the in vitro mechanical properties of this implant have not been reported. To determine the in vitro biomechanical properties of this device, five fresh human cadaver cervical spines were subjected to nondestructive testing in flexion and torsion in three stages: stage 1: intact spine; stage 2: destabilized spine; stage 3: destabilized spine with CSLP. Stage 3 specimens were also subjected to large angular displacement testing to assess the integrity of the fixation. In flexion, mean spinous process displacement was 1.21 mm for stage 1, 3.19 mm for stage 2, and 1.37 mm, for stage 3. Mean torsional stiffness was 2.86 Nm/degree in stage 1, 1.82 Nm/degree in stage 2, and 2.20 Nm/degree in stage 3. Large angular displacement testing in stage 3 resulted in screw loosening from the bone in two specimens; no screw plate loosening occurred. In our severely destabilized in vitro model, the CSLP restored flexion stability but not rotational stability. This suggests that supplemented bracing or fixation may be required to restore torsional stability.

AB - The AO cervical spine locking plate (CSLP) for anterior subaxial fixation was recently received increasing clinical acclaim, yet to date the in vitro mechanical properties of this implant have not been reported. To determine the in vitro biomechanical properties of this device, five fresh human cadaver cervical spines were subjected to nondestructive testing in flexion and torsion in three stages: stage 1: intact spine; stage 2: destabilized spine; stage 3: destabilized spine with CSLP. Stage 3 specimens were also subjected to large angular displacement testing to assess the integrity of the fixation. In flexion, mean spinous process displacement was 1.21 mm for stage 1, 3.19 mm for stage 2, and 1.37 mm, for stage 3. Mean torsional stiffness was 2.86 Nm/degree in stage 1, 1.82 Nm/degree in stage 2, and 2.20 Nm/degree in stage 3. Large angular displacement testing in stage 3 resulted in screw loosening from the bone in two specimens; no screw plate loosening occurred. In our severely destabilized in vitro model, the CSLP restored flexion stability but not rotational stability. This suggests that supplemented bracing or fixation may be required to restore torsional stability.

KW - Anterior fixation

KW - Biomechanics

KW - Cervical spine

KW - Plate

KW - Stability

UR - http://www.scopus.com/inward/record.url?scp=0027503242&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0027503242&partnerID=8YFLogxK

U2 - 10.1007/BF00298363

DO - 10.1007/BF00298363

M3 - Article

VL - 1

SP - 222

EP - 225

JO - European Spine Journal

JF - European Spine Journal

SN - 0940-6719

IS - 4

ER -