Effective ionic dialysance/blood flow rate ratio: An indicator of access recirculation in arteriovenous fistulae

Sumit Mohan, MacHaiah Madhrira, Muhammad Mujtaba, Rajesh Agarwala, Velvie Pogue, Jen Tse Cheng

Research output: Contribution to journalArticle

12 Citations (Scopus)

Abstract

Effective ionic dialysance (EID) is an online measure of hemodialysis (HD) effective urea clearance that is calculated using changes in dialysate sodium conductivity. Effective ionic dialysance is blood flow (Qb) dependent. The presence of significant (≤5%) access recirculation (sAR) during dialysis lowers EID at a given Qb, thereby lowering EID/Qb. We propose using EID/Qb as a useful chairside tool for detection of sAR in arteriovenous fistulae (AVF). Data were collected from 47 patients with AVF (72% men, mean age 49 ± 11.8 years, duration on dialysis 3.78 ± 3.4 years, duration of fistula use 3.35 ± 3.42 years) dialyzed with an high-efficiency dialyzer with a mass transfer area coefficient (KoA) of 1714 ml/min. Effective ionic dialysance were measured at regular intervals by the Gambro Phoenix dialysis system during treatments. The access recirculation (AR) and access blood flow (Qa) were measured using the reference standard saline dilution technique (Transonic HD-02 monitor). Among the 323 HD sessions where Qb, EID, AR, and Qa were available, we identified 17 instances of sAR. The performance of EID/Qb as indicator of sAR was assessed by a receiver operator characteristic (ROC) curve (Stata version 10.1). The area under the ROC curve was 0.935 (95% confidence interval 0.869-1.000), which demonstrated a sensitivity of 76.5% and specificity of 96.4% at an EID/Qb ≤50% with a positive likelihood ratio of 21, negative likelihood ratio of 0.24, positive predictive value of 54.2%, and negative predictive value of 98.7%. We found similar test performance in patients who received HD with dialyzers with smaller surface areas and lower KoAs. The high specificity of EID/Qb makes it an excellent yet simple and early chairside indicator of AVF recirculation.

Original languageEnglish (US)
Pages (from-to)427-433
Number of pages7
JournalASAIO Journal
Volume56
Issue number5
DOIs
StatePublished - Sep 2010
Externally publishedYes

Fingerprint

Dialysis
Arteriovenous Fistula
Renal Dialysis
Blood
Flow rate
Dialysis Solutions
Urea
Indicator Dilution Techniques
Dilution
Mass transfer
Sodium
Fistula
Confidence Intervals
Sensitivity and Specificity
Therapeutics

ASJC Scopus subject areas

  • Biophysics
  • Biomaterials
  • Bioengineering
  • Biomedical Engineering
  • Medicine(all)

Cite this

Effective ionic dialysance/blood flow rate ratio : An indicator of access recirculation in arteriovenous fistulae. / Mohan, Sumit; Madhrira, MacHaiah; Mujtaba, Muhammad; Agarwala, Rajesh; Pogue, Velvie; Cheng, Jen Tse.

In: ASAIO Journal, Vol. 56, No. 5, 09.2010, p. 427-433.

Research output: Contribution to journalArticle

Mohan, Sumit ; Madhrira, MacHaiah ; Mujtaba, Muhammad ; Agarwala, Rajesh ; Pogue, Velvie ; Cheng, Jen Tse. / Effective ionic dialysance/blood flow rate ratio : An indicator of access recirculation in arteriovenous fistulae. In: ASAIO Journal. 2010 ; Vol. 56, No. 5. pp. 427-433.
@article{5b9092f0ac5343729e3be0e1eec1c8c4,
title = "Effective ionic dialysance/blood flow rate ratio: An indicator of access recirculation in arteriovenous fistulae",
abstract = "Effective ionic dialysance (EID) is an online measure of hemodialysis (HD) effective urea clearance that is calculated using changes in dialysate sodium conductivity. Effective ionic dialysance is blood flow (Qb) dependent. The presence of significant (≤5{\%}) access recirculation (sAR) during dialysis lowers EID at a given Qb, thereby lowering EID/Qb. We propose using EID/Qb as a useful chairside tool for detection of sAR in arteriovenous fistulae (AVF). Data were collected from 47 patients with AVF (72{\%} men, mean age 49 ± 11.8 years, duration on dialysis 3.78 ± 3.4 years, duration of fistula use 3.35 ± 3.42 years) dialyzed with an high-efficiency dialyzer with a mass transfer area coefficient (KoA) of 1714 ml/min. Effective ionic dialysance were measured at regular intervals by the Gambro Phoenix dialysis system during treatments. The access recirculation (AR) and access blood flow (Qa) were measured using the reference standard saline dilution technique (Transonic HD-02 monitor). Among the 323 HD sessions where Qb, EID, AR, and Qa were available, we identified 17 instances of sAR. The performance of EID/Qb as indicator of sAR was assessed by a receiver operator characteristic (ROC) curve (Stata version 10.1). The area under the ROC curve was 0.935 (95{\%} confidence interval 0.869-1.000), which demonstrated a sensitivity of 76.5{\%} and specificity of 96.4{\%} at an EID/Qb ≤50{\%} with a positive likelihood ratio of 21, negative likelihood ratio of 0.24, positive predictive value of 54.2{\%}, and negative predictive value of 98.7{\%}. We found similar test performance in patients who received HD with dialyzers with smaller surface areas and lower KoAs. The high specificity of EID/Qb makes it an excellent yet simple and early chairside indicator of AVF recirculation.",
author = "Sumit Mohan and MacHaiah Madhrira and Muhammad Mujtaba and Rajesh Agarwala and Velvie Pogue and Cheng, {Jen Tse}",
year = "2010",
month = "9",
doi = "10.1097/MAT.0b013e3181e743eb",
language = "English (US)",
volume = "56",
pages = "427--433",
journal = "ASAIO Journal",
issn = "1058-2916",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

TY - JOUR

T1 - Effective ionic dialysance/blood flow rate ratio

T2 - An indicator of access recirculation in arteriovenous fistulae

AU - Mohan, Sumit

AU - Madhrira, MacHaiah

AU - Mujtaba, Muhammad

AU - Agarwala, Rajesh

AU - Pogue, Velvie

AU - Cheng, Jen Tse

PY - 2010/9

Y1 - 2010/9

N2 - Effective ionic dialysance (EID) is an online measure of hemodialysis (HD) effective urea clearance that is calculated using changes in dialysate sodium conductivity. Effective ionic dialysance is blood flow (Qb) dependent. The presence of significant (≤5%) access recirculation (sAR) during dialysis lowers EID at a given Qb, thereby lowering EID/Qb. We propose using EID/Qb as a useful chairside tool for detection of sAR in arteriovenous fistulae (AVF). Data were collected from 47 patients with AVF (72% men, mean age 49 ± 11.8 years, duration on dialysis 3.78 ± 3.4 years, duration of fistula use 3.35 ± 3.42 years) dialyzed with an high-efficiency dialyzer with a mass transfer area coefficient (KoA) of 1714 ml/min. Effective ionic dialysance were measured at regular intervals by the Gambro Phoenix dialysis system during treatments. The access recirculation (AR) and access blood flow (Qa) were measured using the reference standard saline dilution technique (Transonic HD-02 monitor). Among the 323 HD sessions where Qb, EID, AR, and Qa were available, we identified 17 instances of sAR. The performance of EID/Qb as indicator of sAR was assessed by a receiver operator characteristic (ROC) curve (Stata version 10.1). The area under the ROC curve was 0.935 (95% confidence interval 0.869-1.000), which demonstrated a sensitivity of 76.5% and specificity of 96.4% at an EID/Qb ≤50% with a positive likelihood ratio of 21, negative likelihood ratio of 0.24, positive predictive value of 54.2%, and negative predictive value of 98.7%. We found similar test performance in patients who received HD with dialyzers with smaller surface areas and lower KoAs. The high specificity of EID/Qb makes it an excellent yet simple and early chairside indicator of AVF recirculation.

AB - Effective ionic dialysance (EID) is an online measure of hemodialysis (HD) effective urea clearance that is calculated using changes in dialysate sodium conductivity. Effective ionic dialysance is blood flow (Qb) dependent. The presence of significant (≤5%) access recirculation (sAR) during dialysis lowers EID at a given Qb, thereby lowering EID/Qb. We propose using EID/Qb as a useful chairside tool for detection of sAR in arteriovenous fistulae (AVF). Data were collected from 47 patients with AVF (72% men, mean age 49 ± 11.8 years, duration on dialysis 3.78 ± 3.4 years, duration of fistula use 3.35 ± 3.42 years) dialyzed with an high-efficiency dialyzer with a mass transfer area coefficient (KoA) of 1714 ml/min. Effective ionic dialysance were measured at regular intervals by the Gambro Phoenix dialysis system during treatments. The access recirculation (AR) and access blood flow (Qa) were measured using the reference standard saline dilution technique (Transonic HD-02 monitor). Among the 323 HD sessions where Qb, EID, AR, and Qa were available, we identified 17 instances of sAR. The performance of EID/Qb as indicator of sAR was assessed by a receiver operator characteristic (ROC) curve (Stata version 10.1). The area under the ROC curve was 0.935 (95% confidence interval 0.869-1.000), which demonstrated a sensitivity of 76.5% and specificity of 96.4% at an EID/Qb ≤50% with a positive likelihood ratio of 21, negative likelihood ratio of 0.24, positive predictive value of 54.2%, and negative predictive value of 98.7%. We found similar test performance in patients who received HD with dialyzers with smaller surface areas and lower KoAs. The high specificity of EID/Qb makes it an excellent yet simple and early chairside indicator of AVF recirculation.

UR - http://www.scopus.com/inward/record.url?scp=77956394883&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=77956394883&partnerID=8YFLogxK

U2 - 10.1097/MAT.0b013e3181e743eb

DO - 10.1097/MAT.0b013e3181e743eb

M3 - Article

C2 - 20577081

AN - SCOPUS:77956394883

VL - 56

SP - 427

EP - 433

JO - ASAIO Journal

JF - ASAIO Journal

SN - 1058-2916

IS - 5

ER -