Evaluation of correlation between Diopsys® NOVATM fixed-luminance flicker ERG and Diagnosys® Espion 2TM flicker ERG parameters

  • Jonathan Regenold
  • , Hien Luong Doan
  • , Hashem Ghoraba
  • , Hassan Khojasteh
  • , Jaclyn Joyce Jaclyn Hwang
  • , Negin Yavari
  • , Amir Akhavanrezayat
  • , Ngoc Trong Tuong Than
  • , Anthony Huy Dinh Le
  • , Muhammad Sohail Halim
  • , Quan Dong Nguyen

Research output: Contribution to journal › Article › peer-review

Abstract

Purpose: Diopsys® NOVA™ is a novel full-field electroretinography (ffERG) device that can make rapid measurements of retinal electrophysiologic function. Diagnosys® Espion 2™ is a clinical gold-standard ERG device. This study aimed to investigate whether light-adapted Diopsys® NOVA™ fixed-luminance flicker ffERG magnitude and implicit time (converted from phase) measurements correlate with light-adapted Diagnosys® Espion 2™ flicker ffERG amplitude and implicit time measurements, respectively. Methods: Twelve patients (22 eyes) with various retinal and uveitic diseases underwent light-adapted Diagnosys® Espion 2™ and Diopsys® NOVA™ fixed-luminance flicker testing. Diopsys® magnitude and implicit time (converted from phase) measurements were compared to Diagnosys® amplitude and implicit time measurements, and a Pearson correlation was used to evaluate any existing correlation. Groups were also compared using generalized estimating equations. Bland–Altman plots were utilized to determine agreement between the comparison groups. Results: Age of patients ranged from 14 to 87 years. 58% (n = 7/12) of patients were female. A significant, positive correlation (r = 0.880, P < 0.001) was observed between magnitude (Diopsys®) and amplitude (Diagnosys®) measurements. Amplitude increases by 6.69 µV for each 1 µV increase in Magnitude (p-value < 0.001). A statistically significant, strong positive correlation was observed between Diopsys® implicit time measurements (converted from phase) and Diagnosys® implicit time measurements (r = 0.814, p-value < 0.001). For each 1 ms increase in Diopsys® implicit time, Diagnosys® implicit time increases by 1.13 ms (p-value < 0.001). Conclusions: There is a statistically significant positive correlation between light-adapted Diopsys® NOVA™ fixed-luminance flicker amplitude and Diagnosys® flicker magnitude values. Additionally, there is a statistically significant positive correlation between Diopsys® NOVA™ fixed-luminance flicker implicit time (converted from phase) and Diagnosys® flicker implicit time values. These results imply that the Diopsys® NOVA™ module, which utilizes the nonstandard shortened International Society for Clinical Electrophysiology of Vision (ISCEV) ERG protocol, can produce reliable light-adapted flicker ffERG measurements.

Original languageEnglish (US)
Pages (from-to)257-266
Number of pages10
JournalDocumenta Ophthalmologica
Volume146
Issue number3
DOIs
StatePublished - Jun 2023
Externally publishedYes

Keywords

  • Diagnosys®
  • Diopsys®
  • Fixed luminance
  • Flicker electroretinography
  • Full-field electroretinography

ASJC Scopus subject areas

  • Ophthalmology
  • Sensory Systems
  • Physiology (medical)

Fingerprint

Dive into the research topics of 'Evaluation of correlation between Diopsys® NOVATM fixed-luminance flicker ERG and Diagnosys® Espion 2TM flicker ERG parameters'. Together they form a unique fingerprint.

Cite this