Abstract
Background: Acute cardiogenic pulmonary edema results from abrupt cardiopulmonary derangements, including arrhythmia or acute valvular dysfunction. Case Summary: An 88-year-old man with nonischemic cardiomyopathy, NYHA functional class III heart failure, left bundle branch block, and severe mitral regurgitation underwent biventricular implantable cardioverter-defibrillator generator replacement. During the procedure, biventricular pacing was interrupted. Within minutes, he developed acute dyspnea, agitation, and hypoxia consistent with flash pulmonary edema. Discussion: Loss of biventricular pacing likely caused ventricular dyssynchrony, worsening mitral regurgitation, and rapidly increasing pulmonary venous pressure. Restoration of pacing improved symptoms. Take-Home Messages: Interruption of biventricular pacing during generator exchange can precipitate flash pulmonary edema, especially in patients with significant mitral regurgitation. Ventricular dyssynchrony from pacing cessation may acutely worsen mitral regurgitation, leading to rapid pulmonary congestion.
| Original language | English (US) |
|---|---|
| Article number | 107942 |
| Journal | JACC: Case Reports |
| Volume | 31 |
| Issue number | 21 |
| DOIs | |
| State | Published - May 27 2026 |
| Externally published | Yes |
Keywords
- acute pulmonary edema
- biventricular pacing
- cardiac resynchronization therapy
- flash pulmonary edema
- generator change
ASJC Scopus subject areas
- Cardiology and Cardiovascular Medicine
Fingerprint
Dive into the research topics of 'Flash Pulmonary Edema From Brief Loss of Biventricular Pacing During CRT-D Generator Exchange'. Together they form a unique fingerprint.Cite this
- APA
- Standard
- Harvard
- Vancouver
- Author
- BIBTEX
- RIS