Measuring Endotracheal Tube Depth by Bedside Ultrasound in Adult Patients in an Intensive Care Unit: A Pilot Study

Fenglin Dong, Canhong Zhu, Huiwen Xu, Jiajia Wang, Yehan Zhu, Qingmin Fan, Jian'an Huang, Wei Lei

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

The aim of the study described here was to evaluate the feasibility and accuracy of measuring endotracheal tube (ETT) depth with ultrasound in adult patients in an intensive care unit (ICU). The distance between the upper margin of the cuff and the upper margin of the aortic arch (Duc-ua) of 67 ICU patients was measured by ultrasound, and the time of measurement was recorded. The level of agreement between the distance between the tip of the ETT and the carina (Dtt-c) measured by ultrasound (U-Dtt-c) and Dtt-c measured by bronchoscopy (B-Dtt-c) was assessed using linear regression and a Bland–Altman plot. There was a significant correlation between B-Dtt-c and U-Dtt-c (r = 0.844, p < 0.001). Also, the Bland–Altman plot revealed strong agreement between B-Dtt-c and U-Dtt-c. The time it took to measure ETT depth by ultrasound was 33.91 ± 5.43 s. In conclusion, bedside ultrasound provides a novel and convenient method for measuring the depth of ETT in ICU patients.

Original languageEnglish (US)
Pages (from-to)1163-1170
Number of pages8
JournalUltrasound in Medicine and Biology
Volume43
Issue number6
DOIs
StatePublished - Jun 2017
Externally publishedYes

Keywords

  • Bronchoscopy
  • Computed tomography scan
  • Critical care
  • Endotracheal intubation
  • Endotracheal tube
  • Point-of-care ultrasound

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Biophysics
  • Acoustics and Ultrasonics

Fingerprint

Dive into the research topics of 'Measuring Endotracheal Tube Depth by Bedside Ultrasound in Adult Patients in an Intensive Care Unit: A Pilot Study'. Together they form a unique fingerprint.

Cite this