TY - JOUR
T1 - Postmarket American Experience with Woven EndoBridge Device
T2 - Adjudicated Multicenter Case Series
AU - Cherian, Jacob
AU - Chen, Stephen R.
AU - Puri, Ajit
AU - Vakharia, Kunal
AU - Levy, Elad
AU - Eshraghi, Sheila
AU - Howard, Brian M.
AU - Tong, Frank C.
AU - Cawley, C. Michael
AU - Gross, Bradley
AU - Alexander, Matthew D.
AU - Grandhi, Ramesh
AU - Srinivasan, Visish M.
AU - Burkhardt, Jan Karl
AU - Johnson, Jeremiah N.
AU - Kan, Peter
N1 - Publisher Copyright:
© 2021 Congress of Neurological Surgeons.
PY - 2021/8/1
Y1 - 2021/8/1
N2 - BACKGROUND: The Woven EndoBridge (WEB) device was granted premarket approval in the United States following results of the Woven EndoBridge Intrasaccular Therapy (WEB-IT) study. WEB-IT reported excellent adequate angiographic occlusion of treated aneurysms with a high safety profile. These results were achieved, however, in the context of a prospective study with strict inclusion criteria and rigorous training support. OBJECTIVE: To review early as-practiced clinical experience with the WEB device in the United States. METHODS: Retrospective review across 6 institutions identified 91 patients undergoing 92 treatment sessions for WEB device placement in treatment of 91 intracranial aneurysms. Details regarding demographics, aneurysm characteristics, treatment considerations, clinical outcomes, and aneurysm occlusion were obtained and analyzed in a multicenter database. Angiograms from the index procedure and follow-up studies were reviewed by a blinded and independent adjudicator. RESULTS: The middle cerebral, anterior communicating, and basilar artery complexes were the commonly treated locations. Eight patients presented with ruptured aneurysms. A mean of 1.2 devices were introduced per case. Technical failure without deployment of a WEB device occurred in 2% (2/92) of sessions. Complete aneurysm occlusion for patients with imaging follow-up was 49% (mean follow-up of 8 mo). Four aneurysms were retreated. 90% of patients had modified Rankin Scale ≤ 2 at last clinical follow-up with no mortalities. CONCLUSION: Immediate postmarket experience with the WEB device, newly introduced at American centers, confirms safe procedural use, but long-term efficacy remains unclear. Early challenges include accurate sizing and device selection.
AB - BACKGROUND: The Woven EndoBridge (WEB) device was granted premarket approval in the United States following results of the Woven EndoBridge Intrasaccular Therapy (WEB-IT) study. WEB-IT reported excellent adequate angiographic occlusion of treated aneurysms with a high safety profile. These results were achieved, however, in the context of a prospective study with strict inclusion criteria and rigorous training support. OBJECTIVE: To review early as-practiced clinical experience with the WEB device in the United States. METHODS: Retrospective review across 6 institutions identified 91 patients undergoing 92 treatment sessions for WEB device placement in treatment of 91 intracranial aneurysms. Details regarding demographics, aneurysm characteristics, treatment considerations, clinical outcomes, and aneurysm occlusion were obtained and analyzed in a multicenter database. Angiograms from the index procedure and follow-up studies were reviewed by a blinded and independent adjudicator. RESULTS: The middle cerebral, anterior communicating, and basilar artery complexes were the commonly treated locations. Eight patients presented with ruptured aneurysms. A mean of 1.2 devices were introduced per case. Technical failure without deployment of a WEB device occurred in 2% (2/92) of sessions. Complete aneurysm occlusion for patients with imaging follow-up was 49% (mean follow-up of 8 mo). Four aneurysms were retreated. 90% of patients had modified Rankin Scale ≤ 2 at last clinical follow-up with no mortalities. CONCLUSION: Immediate postmarket experience with the WEB device, newly introduced at American centers, confirms safe procedural use, but long-term efficacy remains unclear. Early challenges include accurate sizing and device selection.
KW - Brain aneurysm
KW - Intrasaccular flow disruption
KW - WEB
UR - http://www.scopus.com/inward/record.url?scp=85112124955&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85112124955&partnerID=8YFLogxK
U2 - 10.1093/neuros/nyab158
DO - 10.1093/neuros/nyab158
M3 - Article
C2 - 33989398
AN - SCOPUS:85112124955
SN - 0148-396X
VL - 89
SP - 275
EP - 282
JO - Neurosurgery
JF - Neurosurgery
IS - 2
ER -