Quantitative discomanometry

Technique and reproducibility in vitro

Mark A. Fye, Edward Southern, Manohar M. Panjabi, Jacek Cholewicki

Research output: Contribution to journalArticle

16 Citations (Scopus)

Abstract

Quantitative discomanometry is a study of intradiscal pressure and volume measurements during the injection of fluid into an endplate-disc- endplate complex. The purpose of this article was to describe the technique of quantitative discomanometry, determine the reproducibility of the injection technique in a cadaveric thoracolumbar spine, and standardize the technique for future clinical investigations. Nineteen fresh human cadaveric thoracolumbar discs were injected using quantitative discomanometry to determine: (a) the time necessary for a disc to return to a baseline pressure-volume curve, (b) the reproducibility of the technique in vitro, (c) effects of the injection approach and position of the needle in the disc, and (d) effects of the type and length of tubing as well as gauge of spinal needle. A pressure-volume curve was obtained for each disc injection. Reproducibility was measured by nine parameters obtained from each pressure- volume curve: intrinsic pressure, leakage pressure, initial slope, slope between 0-0.1 ml, slope between 1-4 ml, pressure at 2 ml, pressure at 4 ml, maximum pressure, and volume at maximum pressure. The results demonstrated that (a) the injector apparatus was reproducible, (b) the time necessary for a disc to return to a baseline pressure-volume curve was 24 h, (c) the technique using fresh human cadaveric thoracolumbar discs was reproducible, (d) the anterior and posterolateral approaches had similar results if the needle was placed into the center of the nucleus pulposus using radiographic control, and (e) the type and length of tubing, and gauge of needle did not affect the results.

Original languageEnglish (US)
Pages (from-to)335-340
Number of pages6
JournalJournal of Spinal Disorders
Volume11
Issue number4
StatePublished - Aug 28 1998
Externally publishedYes

Fingerprint

Pressure
Needles
Injections
In Vitro Techniques
Spine

Keywords

  • Disc degeneration
  • Discography
  • Discomanometry

ASJC Scopus subject areas

  • Clinical Neurology
  • Surgery

Cite this

Fye, M. A., Southern, E., Panjabi, M. M., & Cholewicki, J. (1998). Quantitative discomanometry: Technique and reproducibility in vitro. Journal of Spinal Disorders, 11(4), 335-340.

Quantitative discomanometry : Technique and reproducibility in vitro. / Fye, Mark A.; Southern, Edward; Panjabi, Manohar M.; Cholewicki, Jacek.

In: Journal of Spinal Disorders, Vol. 11, No. 4, 28.08.1998, p. 335-340.

Research output: Contribution to journalArticle

Fye, MA, Southern, E, Panjabi, MM & Cholewicki, J 1998, 'Quantitative discomanometry: Technique and reproducibility in vitro', Journal of Spinal Disorders, vol. 11, no. 4, pp. 335-340.
Fye, Mark A. ; Southern, Edward ; Panjabi, Manohar M. ; Cholewicki, Jacek. / Quantitative discomanometry : Technique and reproducibility in vitro. In: Journal of Spinal Disorders. 1998 ; Vol. 11, No. 4. pp. 335-340.
@article{0d4f33641d8442cba3e2d95df73b8eeb,
title = "Quantitative discomanometry: Technique and reproducibility in vitro",
abstract = "Quantitative discomanometry is a study of intradiscal pressure and volume measurements during the injection of fluid into an endplate-disc- endplate complex. The purpose of this article was to describe the technique of quantitative discomanometry, determine the reproducibility of the injection technique in a cadaveric thoracolumbar spine, and standardize the technique for future clinical investigations. Nineteen fresh human cadaveric thoracolumbar discs were injected using quantitative discomanometry to determine: (a) the time necessary for a disc to return to a baseline pressure-volume curve, (b) the reproducibility of the technique in vitro, (c) effects of the injection approach and position of the needle in the disc, and (d) effects of the type and length of tubing as well as gauge of spinal needle. A pressure-volume curve was obtained for each disc injection. Reproducibility was measured by nine parameters obtained from each pressure- volume curve: intrinsic pressure, leakage pressure, initial slope, slope between 0-0.1 ml, slope between 1-4 ml, pressure at 2 ml, pressure at 4 ml, maximum pressure, and volume at maximum pressure. The results demonstrated that (a) the injector apparatus was reproducible, (b) the time necessary for a disc to return to a baseline pressure-volume curve was 24 h, (c) the technique using fresh human cadaveric thoracolumbar discs was reproducible, (d) the anterior and posterolateral approaches had similar results if the needle was placed into the center of the nucleus pulposus using radiographic control, and (e) the type and length of tubing, and gauge of needle did not affect the results.",
keywords = "Disc degeneration, Discography, Discomanometry",
author = "Fye, {Mark A.} and Edward Southern and Panjabi, {Manohar M.} and Jacek Cholewicki",
year = "1998",
month = "8",
day = "28",
language = "English (US)",
volume = "11",
pages = "335--340",
journal = "Journal of Spinal Disorders",
issn = "0895-0385",
publisher = "Lippincott Williams and Wilkins Ltd.",
number = "4",

}

TY - JOUR

T1 - Quantitative discomanometry

T2 - Technique and reproducibility in vitro

AU - Fye, Mark A.

AU - Southern, Edward

AU - Panjabi, Manohar M.

AU - Cholewicki, Jacek

PY - 1998/8/28

Y1 - 1998/8/28

N2 - Quantitative discomanometry is a study of intradiscal pressure and volume measurements during the injection of fluid into an endplate-disc- endplate complex. The purpose of this article was to describe the technique of quantitative discomanometry, determine the reproducibility of the injection technique in a cadaveric thoracolumbar spine, and standardize the technique for future clinical investigations. Nineteen fresh human cadaveric thoracolumbar discs were injected using quantitative discomanometry to determine: (a) the time necessary for a disc to return to a baseline pressure-volume curve, (b) the reproducibility of the technique in vitro, (c) effects of the injection approach and position of the needle in the disc, and (d) effects of the type and length of tubing as well as gauge of spinal needle. A pressure-volume curve was obtained for each disc injection. Reproducibility was measured by nine parameters obtained from each pressure- volume curve: intrinsic pressure, leakage pressure, initial slope, slope between 0-0.1 ml, slope between 1-4 ml, pressure at 2 ml, pressure at 4 ml, maximum pressure, and volume at maximum pressure. The results demonstrated that (a) the injector apparatus was reproducible, (b) the time necessary for a disc to return to a baseline pressure-volume curve was 24 h, (c) the technique using fresh human cadaveric thoracolumbar discs was reproducible, (d) the anterior and posterolateral approaches had similar results if the needle was placed into the center of the nucleus pulposus using radiographic control, and (e) the type and length of tubing, and gauge of needle did not affect the results.

AB - Quantitative discomanometry is a study of intradiscal pressure and volume measurements during the injection of fluid into an endplate-disc- endplate complex. The purpose of this article was to describe the technique of quantitative discomanometry, determine the reproducibility of the injection technique in a cadaveric thoracolumbar spine, and standardize the technique for future clinical investigations. Nineteen fresh human cadaveric thoracolumbar discs were injected using quantitative discomanometry to determine: (a) the time necessary for a disc to return to a baseline pressure-volume curve, (b) the reproducibility of the technique in vitro, (c) effects of the injection approach and position of the needle in the disc, and (d) effects of the type and length of tubing as well as gauge of spinal needle. A pressure-volume curve was obtained for each disc injection. Reproducibility was measured by nine parameters obtained from each pressure- volume curve: intrinsic pressure, leakage pressure, initial slope, slope between 0-0.1 ml, slope between 1-4 ml, pressure at 2 ml, pressure at 4 ml, maximum pressure, and volume at maximum pressure. The results demonstrated that (a) the injector apparatus was reproducible, (b) the time necessary for a disc to return to a baseline pressure-volume curve was 24 h, (c) the technique using fresh human cadaveric thoracolumbar discs was reproducible, (d) the anterior and posterolateral approaches had similar results if the needle was placed into the center of the nucleus pulposus using radiographic control, and (e) the type and length of tubing, and gauge of needle did not affect the results.

KW - Disc degeneration

KW - Discography

KW - Discomanometry

UR - http://www.scopus.com/inward/record.url?scp=0031855222&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0031855222&partnerID=8YFLogxK

M3 - Article

VL - 11

SP - 335

EP - 340

JO - Journal of Spinal Disorders

JF - Journal of Spinal Disorders

SN - 0895-0385

IS - 4

ER -