Relationship to drill bit diameter and residual fracture resistance of the distal tibia

Grant R. McChesney, Randal P. Morris, Abeer Al Barghouthi, Francesco Travascio, Loren L. Latta, Ronald W. Lindsey

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Background: The etiology of bone refractures after screw removal can be attributed to residual drill hole defects. This biomechanical study compared the torsional strength of bones containing various sized cortical drill defects in a tibia model. Methods: Bicortical drill hole defects of 3 mm, 4 mm, and 5 mm diameters were tested in 26 composite tibias versus intact controls without a drill defect. Each tibia was secured in alignment with the rotational axis of a materials testing system and the proximal end rotated internally at a rate of 1 deg./s until mechanical failure. Findings: All defect test groups were significantly lower (P < 0.01) in torque-to-failure than the intact group (82.80 ± 3.70 Nm). The 4 mm drill hole group was characterized by a significantly lower (P = 0.021) torque-to-failure (51.00 ± 3.27 Nm) when compared to the 3 mm drill hole (59.00 ± 5.48 Nm) group, but not different than the 5 mm hole group (55.71 ± 5.71 Nm). All bones failed through spiral fractures, bones with defects also exhibited posterior butterfly fragments. Interpretation: All the tested drill hole sizes in this study significantly reduced the torque-to-failure from intact by a range of 28.4% to 38.4%, in agreement with previous similar studies. The 5 mm drill hole represented a 22.7% diameter defect, the 4 mm drill hole a 18.2% diameter defect, and the 3 mm drill hole a 13.6% diameter defect. Clinicians should be cognizant of this diminution of long bone strength after a residual bone defect in their creation and management of patient rehabilitation programs.

Original languageEnglish (US)
Article number105686
JournalClinical Biomechanics
Volume97
DOIs
StatePublished - Jul 2022
Externally publishedYes

Keywords

  • Composite tibia
  • Residual defects
  • Screw hole
  • Torsional strength
  • drill hole

ASJC Scopus subject areas

  • Biophysics
  • Orthopedics and Sports Medicine

Fingerprint

Dive into the research topics of 'Relationship to drill bit diameter and residual fracture resistance of the distal tibia'. Together they form a unique fingerprint.

Cite this