Shielding evaluation for IMRT implementation in an existing accelerator vault.

R. A. Price, O. Chibani, C. M. Ma

Research output: Contribution to journalArticlepeer-review

14 Scopus citations

Abstract

A formalism is developed for evaluating the shielding in an existing vault to be used for IMRT. Existing exposure rate measurements are utilized as well as a newly developed effective modulation scaling factor. Examples are given for vaults housing 6, 10 and 18 MV linear accelerators. The use of an 18 MV Siemens linear accelerator is evaluated for IMRT delivery with respect to neutron production and the effects on individual patients. A modified modulation scaling factor is developed and the risk of the incurrence of fatal secondary malignancies is estimated. The difference in neutron production between 18 MV Varian and Siemens accelerators is estimated using Monte Carlo results. The neutron production from the Siemens accelerator is found to be approximately 4 times less than that of the Varian accelerator resulting in a risk of fatal secondary malignancy occurrence of approximately 1.6% when using the SMLC delivery technique and our measured modulation scaling factors. This compares with a previously published value of 1.6% for routine 3D CRT delivery on the Varian accelerator. (c) 2003 American College of Medical Physics.

Original languageEnglish (US)
Pages (from-to)231-238
Number of pages8
JournalJournal of applied clinical medical physics / American College of Medical Physics
Volume4
Issue number3
DOIs
StatePublished - 2003
Externally publishedYes

ASJC Scopus subject areas

  • Radiation
  • Instrumentation
  • Radiology Nuclear Medicine and imaging

Fingerprint

Dive into the research topics of 'Shielding evaluation for IMRT implementation in an existing accelerator vault.'. Together they form a unique fingerprint.

Cite this