The added value of volume-of-interest C-arm CT imaging during endovascular treatment of intracranial aneurysms

G. Chintalapani, P. Chinnadurai, A. Maier, Y. Xia, S. Bauer, Hashem Shaltoni, H. Morsi, M. E. Mawad

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

BACKGROUND AND PURPOSE: Successful endovascular treatment of intracranial aneurysms requires understanding the exact relationship of implanted devices to the aneurysm, parent artery, and other branch vessels during the treatment. Intraprocedural C-arm CT imaging has been shown to provide such information. However, its repeated use is limited due to increasing radiation exposure to the patient. The goal of this study was to evaluate a new volume-of-interest C-arm CT imaging technique, which would provide device-specific information through multiple 3D acquisitions of only the region of interest, thus reducing cumulative radiation exposure to the patient. MATERIALS AND METHODS: VOI C-arm CT images were obtained in 28 patients undergoing endovascular treatment of intracranial aneurysms. VOI images were acquired with the x-ray source collimated around the deployed device, both horizontally and vertically. The images were reconstructed by using a novel prototype robust reconstruction algorithm to minimize truncation artifacts from double collimation. The reconstruction accuracy of VOI C-arm CT images was assessed quantitatively by comparing them with the full-head noncollimated images. RESULTS: Quantitative analysis showed that the quality of VOI C-arm CT images is comparable with that of the standard Feldkamp, Davis, and Kress reconstruction of noncollimated C-arm CT images (correlation coefficient = 0.96 and structural similarity index = 0.92). Furthermore, 91.5% reduction in dose-area product was achieved with VOI imaging compared with the full-head acquisition. CONCLUSIONS: VOI imaging allows multiple 3D C-arm CT acquisitions and provides information related to device expansion, parent wall apposition, and neck coverage during the procedure, with very low additional radiation exposure to the patient.

Original languageEnglish (US)
Pages (from-to)660-666
Number of pages7
JournalAmerican Journal of Neuroradiology
Volume37
Issue number4
DOIs
StatePublished - Apr 1 2016
Externally publishedYes

Fingerprint

Intracranial Aneurysm
Arm
Equipment and Supplies
Therapeutics
Head
Artifacts
Aneurysm
Neck
Arteries
X-Rays
Radiation Exposure

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Clinical Neurology

Cite this

The added value of volume-of-interest C-arm CT imaging during endovascular treatment of intracranial aneurysms. / Chintalapani, G.; Chinnadurai, P.; Maier, A.; Xia, Y.; Bauer, S.; Shaltoni, Hashem; Morsi, H.; Mawad, M. E.

In: American Journal of Neuroradiology, Vol. 37, No. 4, 01.04.2016, p. 660-666.

Research output: Contribution to journalArticle

Chintalapani, G. ; Chinnadurai, P. ; Maier, A. ; Xia, Y. ; Bauer, S. ; Shaltoni, Hashem ; Morsi, H. ; Mawad, M. E. / The added value of volume-of-interest C-arm CT imaging during endovascular treatment of intracranial aneurysms. In: American Journal of Neuroradiology. 2016 ; Vol. 37, No. 4. pp. 660-666.
@article{106fd113dbeb4faabc6217e9d9ebe162,
title = "The added value of volume-of-interest C-arm CT imaging during endovascular treatment of intracranial aneurysms",
abstract = "BACKGROUND AND PURPOSE: Successful endovascular treatment of intracranial aneurysms requires understanding the exact relationship of implanted devices to the aneurysm, parent artery, and other branch vessels during the treatment. Intraprocedural C-arm CT imaging has been shown to provide such information. However, its repeated use is limited due to increasing radiation exposure to the patient. The goal of this study was to evaluate a new volume-of-interest C-arm CT imaging technique, which would provide device-specific information through multiple 3D acquisitions of only the region of interest, thus reducing cumulative radiation exposure to the patient. MATERIALS AND METHODS: VOI C-arm CT images were obtained in 28 patients undergoing endovascular treatment of intracranial aneurysms. VOI images were acquired with the x-ray source collimated around the deployed device, both horizontally and vertically. The images were reconstructed by using a novel prototype robust reconstruction algorithm to minimize truncation artifacts from double collimation. The reconstruction accuracy of VOI C-arm CT images was assessed quantitatively by comparing them with the full-head noncollimated images. RESULTS: Quantitative analysis showed that the quality of VOI C-arm CT images is comparable with that of the standard Feldkamp, Davis, and Kress reconstruction of noncollimated C-arm CT images (correlation coefficient = 0.96 and structural similarity index = 0.92). Furthermore, 91.5{\%} reduction in dose-area product was achieved with VOI imaging compared with the full-head acquisition. CONCLUSIONS: VOI imaging allows multiple 3D C-arm CT acquisitions and provides information related to device expansion, parent wall apposition, and neck coverage during the procedure, with very low additional radiation exposure to the patient.",
author = "G. Chintalapani and P. Chinnadurai and A. Maier and Y. Xia and S. Bauer and Hashem Shaltoni and H. Morsi and Mawad, {M. E.}",
year = "2016",
month = "4",
day = "1",
doi = "10.3174/ajnr.A4605",
language = "English (US)",
volume = "37",
pages = "660--666",
journal = "American Journal of Neuroradiology",
issn = "0195-6108",
publisher = "American Society of Neuroradiology",
number = "4",

}

TY - JOUR

T1 - The added value of volume-of-interest C-arm CT imaging during endovascular treatment of intracranial aneurysms

AU - Chintalapani, G.

AU - Chinnadurai, P.

AU - Maier, A.

AU - Xia, Y.

AU - Bauer, S.

AU - Shaltoni, Hashem

AU - Morsi, H.

AU - Mawad, M. E.

PY - 2016/4/1

Y1 - 2016/4/1

N2 - BACKGROUND AND PURPOSE: Successful endovascular treatment of intracranial aneurysms requires understanding the exact relationship of implanted devices to the aneurysm, parent artery, and other branch vessels during the treatment. Intraprocedural C-arm CT imaging has been shown to provide such information. However, its repeated use is limited due to increasing radiation exposure to the patient. The goal of this study was to evaluate a new volume-of-interest C-arm CT imaging technique, which would provide device-specific information through multiple 3D acquisitions of only the region of interest, thus reducing cumulative radiation exposure to the patient. MATERIALS AND METHODS: VOI C-arm CT images were obtained in 28 patients undergoing endovascular treatment of intracranial aneurysms. VOI images were acquired with the x-ray source collimated around the deployed device, both horizontally and vertically. The images were reconstructed by using a novel prototype robust reconstruction algorithm to minimize truncation artifacts from double collimation. The reconstruction accuracy of VOI C-arm CT images was assessed quantitatively by comparing them with the full-head noncollimated images. RESULTS: Quantitative analysis showed that the quality of VOI C-arm CT images is comparable with that of the standard Feldkamp, Davis, and Kress reconstruction of noncollimated C-arm CT images (correlation coefficient = 0.96 and structural similarity index = 0.92). Furthermore, 91.5% reduction in dose-area product was achieved with VOI imaging compared with the full-head acquisition. CONCLUSIONS: VOI imaging allows multiple 3D C-arm CT acquisitions and provides information related to device expansion, parent wall apposition, and neck coverage during the procedure, with very low additional radiation exposure to the patient.

AB - BACKGROUND AND PURPOSE: Successful endovascular treatment of intracranial aneurysms requires understanding the exact relationship of implanted devices to the aneurysm, parent artery, and other branch vessels during the treatment. Intraprocedural C-arm CT imaging has been shown to provide such information. However, its repeated use is limited due to increasing radiation exposure to the patient. The goal of this study was to evaluate a new volume-of-interest C-arm CT imaging technique, which would provide device-specific information through multiple 3D acquisitions of only the region of interest, thus reducing cumulative radiation exposure to the patient. MATERIALS AND METHODS: VOI C-arm CT images were obtained in 28 patients undergoing endovascular treatment of intracranial aneurysms. VOI images were acquired with the x-ray source collimated around the deployed device, both horizontally and vertically. The images were reconstructed by using a novel prototype robust reconstruction algorithm to minimize truncation artifacts from double collimation. The reconstruction accuracy of VOI C-arm CT images was assessed quantitatively by comparing them with the full-head noncollimated images. RESULTS: Quantitative analysis showed that the quality of VOI C-arm CT images is comparable with that of the standard Feldkamp, Davis, and Kress reconstruction of noncollimated C-arm CT images (correlation coefficient = 0.96 and structural similarity index = 0.92). Furthermore, 91.5% reduction in dose-area product was achieved with VOI imaging compared with the full-head acquisition. CONCLUSIONS: VOI imaging allows multiple 3D C-arm CT acquisitions and provides information related to device expansion, parent wall apposition, and neck coverage during the procedure, with very low additional radiation exposure to the patient.

UR - http://www.scopus.com/inward/record.url?scp=84963904735&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84963904735&partnerID=8YFLogxK

U2 - 10.3174/ajnr.A4605

DO - 10.3174/ajnr.A4605

M3 - Article

VL - 37

SP - 660

EP - 666

JO - American Journal of Neuroradiology

JF - American Journal of Neuroradiology

SN - 0195-6108

IS - 4

ER -